OBJECTIVES: To assess quantitatively the combined effect of exomass-related- and motion artefacts on voxel value parameters in cone beam CT (CBCT). METHODS: A cylindrical phantom was manufactured, containing 21 tubes filled with a radiopaque solution, allowing the inclusion of three titanium implants in the periphery to induce exomass-related artefacts. The phantom was mounted on a robot simulating 0.75-, 1.5-, and 3 mm movements (nodding/lateral rotation/tremor). CBCT images with/without exomass and with/without movements were acquired in duplicate in three units: Cranex 3Dx, Orthophos SL-3D, and X1 (with motion-artefact correction). A cylindrical volume of interest was defined in each tube and voxel value mean and standard deviation were assessed. For each CBCT volume, the 21 mean voxel values were averaged providing the overall mean voxel value (MVV), and the standard deviation (among the 21 values) was calculated providing overall voxel value inhomogeneity (VVI). The standard deviation from each of the 21 volumes-of-interest were averaged, providing overall image noise (IN). The effect of the diverse tested situations was inferred from a repeated-measures analysis of variance, followed by Sidak's test (α = 0.05). RESULTS: Overall, images acquired with exomass showed significantly (p ≤ 0.05) lower MVV, and higher VVI and IN. Motion artefacts aggravated exomass-related alterations. MVV and VVI were mostly affected by 3 mm nodding movements. Motion-artefact correction eliminated the deleterious effect of movement. CONCLUSIONS: CBCT voxel-value parameters are altered by exomass-related artefacts, and this finding is aggravated in the presence of motion artefacts. Motion-artefact correction effectively eliminated the deleterious impact of movement.


    Access

    Download


    Export, share and cite



    Title :

    Objective assessment of the combined effect of exomass-related- and motion artefacts in cone beam CT



    Publication date :

    2021-01-01


    Remarks:

    Oliveira , M L , Candemil , A P , Freitas , D Q , Haiter-Neto , F , Wenzel , A & Spin-Neto , R 2021 , ' Objective assessment of the combined effect of exomass-related- and motion artefacts in cone beam CT ' , Dento maxillo facial radiology , vol. 50 , no. 1 , 20200255 . https://doi.org/10.1259/dmfr.20200255



    Type of media :

    Article (Journal)


    Type of material :

    Electronic Resource


    Language :

    English



    Classification :

    DDC:    629



    Reduction of motion artefacts in on-board cone beam CT by warping of projection images

    Marchant, T. E. / Price, G. J. / Matuszewski, B. J. et al. | BASE | 2011

    Free access

    Surgical Video Motion Magnification with Suppression of Instrument Artefacts

    Janatka, M / Marcus, HJ / Dorward, NL et al. | BASE | 2020

    Free access

    Suppressing motion artefacts in MRI using an Inception‐ResNet network with motion simulation augmentation

    Pawar, Kamlesh / Chen, Zhaolin / Shah, N. J. et al. | BASE | 2022

    Free access

    A Motion Compensation Approach for Oral and Maxillofacial Cone-beam Imaging

    Sun, Tao / Jacobs, Reinhilde / Pauwels, Ruben et al. | BASE | 2020

    Free access

    Surrogate-Driven Motion Model for Motion Compensated Cone-beam CT Reconstruction using Unsorted Projection Data

    Huang, Yuliang / Thielemans, Kris / McClelland, Jamie R | BASE | 2023

    Free access